SER-9513 Tinytag Transit Service Kit
  • SER-9513 Tinytag Transit Service Kit

SER-9513

Quick Overview

A service kit for the Tinytag Transit (TG-3080). This kit contains a replacement battery and a service instruction sheet.

Details

SER-9513 Tinytag Transit (TG-3080) Service Kit

For use with logger models:
TG-3080.

1. Remove the data logger’s lid and ‘O’ ring, and gently slide the battery out of its holder.
2. Insert the new battery, observing the correct polarity, and taking care not to touch the unit’s circuit board.
3. For data loggers below SN 251690: The unit’s green LED should come on for four seconds and then go out. If this does not happen, remove the battery for 30 seconds and try again.
    For data loggers above SN 251690: Both LEDs will flash briefly.
4. Re-fit the ’O’ ring to secure the battery in its holder.
5. Place the data logger on its inductive pad and check that Tinytag Explorer can communicate with it.
6. Refit the lid.

The TG-3080 has been replaced with the new Tinytag Transit 2 data logger. A link below is provided to the new Tinytag Transit 2 data logger page.

Availability: Stock Item