Description
The SPA-PCB-PE is the Filter Carrier and PE Filter used on the standard HC2A-S HygroClip probe, but can be used with most of the HC2A-S range of probes, except the HC2A-SM or Industrial range of HygroClip probes. SPA-PCB-PE is suitable for most applications and provides protection from atmospheric dust and moisture. A Stevenson Screen is advised for outdoor use to provide additional protection from wind and precipitation.
Specifications
Carrier Only | With PE Filter | With PTFE Filter | |
Part Number | SPA-PCB | SPA-PCB-PE | SPA-PCB-PTFE |
Filter | ![]() |
![]() |
![]() |
Material | PC / 1.4301 | Polyethylene | PTFE |
Pore Size | – | 40-50µm | 10µm |
Temperature Range | -50…100°C (-58…212F) | -50…100°C (-58…212F) | -80…100°C (-112…212F) |
Response Time %RH (*1) | 12 Seconds | 15 Seconds | 18 Seconds |
Response Time °C (*2) | 80 Seconds | 180 Seconds | 170 Seconds |
Waterproof (Dripping 50mm) |
No | Yes | Yes |
Max. Wind Speed | 5 m/s | 50 m/s | 50 m/s |
Suitable For Fine Dust | No | >10 µm | Yes |
General Information | Fast Response Time | Standard Filter | High Chemical Resistance, Fine Dust Resistant, High Temperature |
Application | Sterilisation, H2O2, Handheld Devices | HVAC, Office, Storage Rooms, Laboratory, Clean Rooms | Cleanrooms, Medical, Injection Chamber, Meal Manufacturing |
*1 T63: 100%RH → 30%RH
*2 PC carrier: 70 °C → 23 °C
.