Description
Choosing the correct transducer is critical to matching your specific material type, thickness range, geometry, and surface temperature.
Detailed Probe Profiles & Technical Data
N05 / N05/90° — The Standard Utility Probe
- Description: The default workhorse dual-element transducer supplied with most MT-series units. Balanced sensitivity and penetration for general engineering inspections.
- Best Used For: Structural steel, raw plates, large-diameter tanks, and machined components.
- Limitations: Avoid using on active steam lines or pipe walls thinner than 1.2 mm.
N07 — The Micro-Bore & Thin-Wall Probe
- Description: High-frequency, small-footprint transducer designed to focus acoustic energy into thin boundaries.
- Best Used For: Small-diameter tubing, curved elbows, hydraulic lines, and copper heat exchangers.
- Technical Tip: Requires extra couplant and steady, vertical pressure due to the reduced surface contact area.
HT5 — High-Temperature Transducer
- Description: Engineered with specialised thermal-insulating delay lines and heavy-duty casing to withstand extreme heat cycles.
- Best Used For: In-service oil and gas pipelines, refinery vessels, and steam pipe monitoring.
- Crucial Operation Rule:
Note on High-Temp Testing: Limit physical probe contact to under 5 seconds when measuring surfaces up to 300°C. Allow the probe to air-cool for at least 15 seconds between measurements to avoid internal crystal damage. Use specialised high-temperature silicone couplant.
N02 — Thick & High-Attenuation Probe
- Description: Low-frequency design with massive penetration capabilities. Low frequencies suffer less scattering in coarse, non-homogeneous structures.
- Best Used For: Grey cast iron, ductile iron, fibreglass, composite structures, and heavy rubber linings.
P5EE — Echo-to-Echo Thru-Paint Transducer
- Description: Developed strictly for gauges featuring Echo-to-Echo firmware. It filters out the thickness of protective surface coatings (paints, epoxies) to read only the underlying metal substrate.
- Best Used For: Painted storage tanks, coated structural columns, and marine hulls.
- Note: Not compatible with base MT150 / MT160 units. Requires MT180 or MT190.
Quick-Reference Probe Compatibility & Specification
| Probe Model | Frequency | Contact Diameter | Measurement Range (Steel) | Temperature Range | Primary Application | Compatible MT Gauges |
|---|---|---|---|---|---|---|
N05![]() |
5.0 MHz | 10 mm | 1.2 mm to 230.0 mm | -10°C to 60°C | Standard/General-purpose flat surfaces | MT150, MT160, MT180, MT190, MT200, MT600, MT660 |
N05 / 90°![]() |
5.0 MHz | 10 mm | 1.2 mm to 230.0 mm | -10°C to 60°C | Right-angle cable exit for tight physical spaces | MT150, MT160, MT180, MT190, MT200, MT600, MT660 |
N07![]() |
7.0 MHz | 6 mm | 0.75 mm to 80.0 mm | -10°C to 60°C | Thin-walled tubes, small diameters, tight curves | MT150, MT160, MT180, MT190, MT200, MT600, MT660 |
HT5![]() |
5.0 MHz | 12 mm | 3.0 mm to 200.0 mm | Up to 300°C | High-temperature pipes, active process vessels | MT150, MT160, MT180, MT190, MT200, MT600, MT660 |
N02![]() |
2.5 MHz | 14 mm | 3.0 mm to 300.0 mm | -10°C to 60°C | Attenuative, coarse-grained, or cast materials | MT150, MT160, MT180, MT190, MT200, MT600, MT660 |
P5EE![]() |
5.0 MHz | 10 mm | 3.0 mm to 150.0 mm (EE) 2.0 mm to 600.0 mm (PE) |
-10°C to 60°C | Echo-to-Echo (EE): Thru-paint & coatings | MT180, MT190, MT600, MT660 |
Please confirm the meter the probes will be used with to ensure the correct probes are supplied.