omni instruments | sensors and data acquisition equipment

Sales & Technical Support

Edit Content

IF-96012

IF96012 Plug in RS485 Modbus/Jbus communication module for the Nemo 96HD Multi-function power meter. Internal memory for up to 5,760 hours of recorded data. 3 screw terminal connection. Fits into slot A of the Nemo 96HD/HD+/HDLe Power Meters, Max 1 per meter.

£184.63

Need help finding the right product?

Downloads

NEMO 96HD Modules

Description

IF-96012: RS485 Interface for Nemo 96HD/HD+/HDLe Series

The IF-96012 module, used in conjunctions with Nemo 96HD/HD+/HDLe series, allows all the data relevant to measurements and configuration parameters to be retrieved via an RS485 Interface. Thanks to its internal memory, it stores energy counts and main measured parameters.

Memory

Measurements are saved in a sliding-window memory, where the oldest data is overwritten by the newest. The memory allows reading previous data, with a time depth that varies depending on the saved data quantity and the saving interval.

Programmable Parameters – Memory

  • Clock: hour, minutes, seconds
  • Date: day, month, year
  • Daylight saving time: start and end date and time
  • Time interval between data backups: 2-5-10-30-60 seconds – 2-5-10 minutes
  • Saved data: type 0/1/2/3
  • Reset saved data

RS485 Communication Specifications:

  • Isolation: Galvanically insulated from measurement inputs and auxiliary power.
  • Standard: 3-Wire RS485.
  • Transmission: Serial asynchronous.
  • Protocol: JBUS/MODBUS.
  • Bit number: 8.
  • Stop bit: 1.
  • Response time: Required response time to request: ≤200ms.
  • Meters that can be connected on the bus: 32 (up to 255 with RS485 repeater.
  • Max distance from supervisor: 1200m.

Programmable Parameters – RS485

  • Address: 1…255
  • Baud rate: 4.800-9.600-19.200-38.400 bit/s
  • Parity bit: none – even – odd

Auxiliary Supply

  • Power Consumption: Rated burden: ≤5VA.

Insulation (EN61010):

  • Testing Voltage: A.C. voltage test: 2kV r.m.s. value at 50Hz for 1 minute.
  • Considered Circuits: Measurement input, auxiliary supply, RS485 communication.

Housing:

  • Type: Module with a connector for the Nemo 96HD/HD+/HDLe.
  • Maximum Depth: 81mm (including the module).
  • Connections: Screw terminals.
  • Terminals Range: Rigid cable max 4.5mm², flexible cable max 2.4mm²
  • Material: Self-extinguishing polycarbonate.
  • Weight: 50 grams.

Firmware Version:

The table indicates compatible meters for the IF96012 module, along with position constraints and the maximum number of modules usable. It also shows the firmware version of the meter necessary to support the additional module’s function.

Code Description Multifunction Model N. Max. Position Firmware
A B C D
 

IF96012

 

Interface Memory + RS485

Nemo 96HD 1 2.06
Nemo 96HD+ 1 2.06
Nemo 96HDLe 1 2.50
Nemo 96HDLe Rogowski 1 1.00

The firmware version of the meter required to support the module’s function is indicated in the documentation. Firmware can be updated in the field using an IF-96001 (RS485) or IF-96002 (RS232), or IF-96015 (Ethernet) communication module, along with a PC and the necessary software.

We are here to help

Need help with your products or have a question for us on our custom systems? Use the form below to send us a message and our experts will be in touch to help.

Enquire now