Description
Traditionally, DO measurement in portable field equipment has been done using membrane covered detectors known as Clark Cells. This type of cell suffers from problems including membrane fouling, calibration instability and worst of all, oxygen consumption. During measurement, a Clark Cell will consume oxygen making it necessary to have a constant flow of water over the cell.
Optical technology eliminates all these problems allowing high precision, membrane-free, long-term stability along with infrequent calibration and immunity to fouling by sulphides and other gases.
The Aquaread AquaPlus is the only Optical DO system that measures salinity directly. This allows for automatic salinity compensation giving you the highest accuracy in any type of water.
The Aquaread AquaPlus works on the principle of Dynamic Luminescence Quenching. A gas-permea-ble material known as a luminophore is excited with short bursts of blue light, which causes mole-cules in the luminophore to emit red photons. By measuring the delay of the returned red photons with respect to the blue excitation, the level of dissolved oxygen present can be determined.
AquaPlus with the sleeve removed reveals the combined dissolved oxygen, conductivity & temperature sensors. The end cap is replaceable however you can expect more than 2 years life from one cap
Specifications
Protection Class: | IP68 (Permanent immersion to 30m, 1 week to 100m) |
Immersion Depth: | Min 75mm, Max 100m |
Operating Temperature: | -5 to +70°C |
Dimensions (L x Dia): | 250mm x 24mm |
Weight: | 400g |