AP-2000 Ion Selective Electrodes

Aquaprobe Ion Selective Electrodes (ISE Sensors) for the AP-2000 digital water quality sensor. These advanced ion selective electrodes for monitoring specific ions in the water. They are incredibly practical, allowing for easy, tool-free removal and replacement in the field. They’re designed specifically for seamless integration with the AP-2000 sensor, to provide a additional parameters to your water quality monitor sensor.

Need help finding the right product?

Description

The AP-2000 water quality probe is a multi-parameter sensor and comes pre-fitted with probes covering pH, ORP, Conductivity, TDS, SSG, Resistivity, Salinity, Dissolved Oxygen and Temperature probes, the AP-2000-D add a depth sensor. Each probe comes with two empty sockets to add an Optical, or Ion Selective Electrode (ISE) sensors. This page covers the 5 Ion Selective Sensors that can be fitted to Aux Port 2 on the AP-2000 probes to add one more parameter to the probe. Additional 6 parameters covered are: Ammonium; Ammonia; Chloride; Fluoride; Nitrate; or Calcium.

The table below covers the available parameters, with links to get more details on the parameter for each probe.

Ion Selective Electrode (ISE) Options

Ammonium Range 0 – 9,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.0-99.9 mg/L 100 – 8,999 mg/L
Accuracy ±10% of reading or 2ppm (whichever is greater)
Ammonia* Range 0 – 9,000 mg/L (ppm)
Resolution 2 Auto-range scales: 0.0-99.9 mg/L 100 – 8,999 mg/L
Accuracy ±10% of reading or 2ppm (whichever is greater)
Chloride Range 0 – 20,000 mg/L (ppm)
Resolution 2 Auto-range scales: 0.0-99.9 mg/L 100 – 19,999 mg/L
Accuracy ±10% of reading or 2ppm (whichever is greater)
Fluoride Range 0 – 1,000 mg/L (ppm)
Resolution 2 Auto-range scales: 0.0-99.9 mg/L 100 – 999 mg/L
Accuracy ±10% of reading or 2ppm (whichever is greater)
Nitrate Range 0 – 30,000 mg/L (ppm)
Resolution 2 Auto-range scales: 0.0-99.9 mg/L 100 – 29,999 mg/L
Accuracy ±10% of reading or 2ppm (whichever is greater)
Calcium Range 0 – 2,000 mg/L (ppm)
Resolution 2 Auto-range scales: 0.0-99.9 mg/L 100 – 1,999 mg/L
Accuracy ±10% of reading or 2ppm (whichever is greater)

* Ammonium electrode required. Readings calculated from ammonium, pH and temperature values.

The accuracy figures quoted above represent the equipment’s capability at the calibration points at 25°C. These figures do not take into account errors introduced by variations in the accuracy of calibration solutions and errors beyond the control of the manufacturer that may be introduced by environmental conditions in the field.

The AP-2000 Aux Ports

The AP-2000 probe has two available Aux ports. Aux Port 1 can be fitted with either an Optical or ISE sensor. Both ports can be used with any of the five ISE sensors to add any of the six parameters. The AP-2000 Optical Electrodes can be found on the link below.

AP-2000 Optical Electrodes

Product

Ref

Price

Product Description

2000-AMM

£470.00

AP-2000 Ammonium Electrode with 2 Auto-Range scales 0-99.9 mg/L (ppm) or 100-8,999 mg/L (ppm). Accuracy ±10% of readings or 2ppm (whichever is greater). Probe will also monitor Ammonia when used in combination with pH and temperature readings on the AP-2000 Aquaprobe sensor

2000-CAL

£470.00

AP-2000 Calcium Electrode with 2 Auto-Range scales 0-99.9 mg/L (ppm) or 100-1,999 mg/L (ppm). Accuracy ±10% of readings or 2ppm (whichever is greater). Minimum Level of Detection 0.05 ppm. The 2000-CAL is designed for use with the AP-2000 Aquaprobe sensor with pH probe.

2000-CHL

£470.00

AP-2000 Chloride Electrode with 2 Auto-Range scales 0-99.9 mg/L (ppm) or 100-8,999 mg/L (ppm). Accuracy ±10% of readings or 2ppm (whichever is greater). The 2000-CHL is designed for use with the AP-2000 Aquaprobe sensor

2000-FLU

£470.00

AP-2000 Fluoride Electrode with 2 Auto-Range scales 0-99.9 mg/L (ppm) or 100-999 mg/L (ppm). Accuracy ±10% of readings or 2ppm (whichever is greater). Minimum Level of Detection 0.05 ppm. The 2000-FLU is designed for use with the AP-2000 Aquaprobe sensor with pH probe. Readings may be affected by Hydroxide ions.

2000-NIT

£470.00

AP-2000 Nitrate Electrode with 2 Auto-Range scales 0-99.9 mg/L (ppm) or 100-29,999 mg/L (ppm). Accuracy ±10% of readings or 2ppm (whichever is greater). Minimum Level of Detection 0.5 ppm. The 2000-NIT is designed for use with the AP-2000 Aquaprobe sensor with pH probe.