17124975-G

Replacement MX6 Sensor for Carbon Monoxide/Hydrogen Null (CO/H2 null). Range 0 to 1,000ppm, Resolution 1ppm. Response time 9 seconds (T50). This cell will provide a more accurate reading in areas where H2 is present, than the standard CO cell.

£556.00

Need help finding the right product?

Description

Carbon Monoxide (CO) is a colourless, odourless, tasteless, and non-irritating gas that is highly toxic. It’s often referred to as the “silent killer” because it’s impossible to detect with human senses, and its presence can lead to severe health effects or death without warning.

Key Properties

  • Colourless, Odourless, Tasteless: Undetectable by human senses
  • Highly Toxic: Interferes with the blood’s ability to carry oxygen
  • Slightly Lighter Than Air: It mixes readily with air, but can still accumulate in enclosed spaces if not properly ventilated
  • Flammable: Can ignite and form explosive mixtures in high concentrations

 

Replacement Sensor, Part No. 17124975-G, Properties

Gas Detected
(Abbreviation)
Sensor Type Measurement Response time
(Seconds)
Calibration Gas
& Concentration
Range Resolution T50 T90
Carbon Monoxide/Hydrogen Low
(CO/H2 Low)
Electrochemical 0 to 1.000ppm 1.0 ppm 9 20 CO
100 ppm

This replacement cell is designed for gas detection where there may be H2 present in the air. The H2 can create a false reading for CO, as shown in the table below, and 17124975-G cell has a much lower sensitivity to H2 than the standard 17124975-1 cell. A 100ppm level of H2 would create a false reading of 22ppm in CO, compared to just 3ppm with the appropriate CO/H2 Low cell.

Toxic Gas Sensor Cross-sensitivity Table

Target Gas Sensor
CO CO
(H2 Low)
H2S SO2 NO2 Cl2 ClO2 HCN HCl PH3 NO H2 NH3
CO 100 100 1 1 0 0 0 0 0 0 0 20 0
H2S 5 5 100 1 -40 -3 -25 10 300 25 10 20 25
SO2 0 5 5 100 0 0 0 40 0 0 -40
NO2 -5 5 -25 -165 100 45 -70 30 0 -10
Cl2 -10 0 -20 -25 10 100 60 -20 6 -20 0 0 -50
ClO2 20 100
HCN 15 50 1 0 0 100 35 1 0 30 5
HCl 3 5 0 2 0 0 100 0 15 0 0
PH3 -100 425 300 100
NO 25 40 -0.2 1 5 -5 100 30 0
H2 22 3 0.08 0.5 0 0 0 0 0 0 0 100 0
NH3 0 0 0 0 0 0 0 0 0 0 0 0 100

The table above reflects the percentage response provided by the sensor (top row) when exposed to a known concentration of the target gas (column 1).
The numbers were measured under these environmental conditions: 20 °C (68 °F), 50% RH and 1 atm.
The specified cross-interference numbers apply to new sensors only and may vary with time as well as from sensor to sensor.
“-” means no data available.
This table is given as a reference only and is subject to change.

 