Replacement MX6 Sensor for Sulphur Dioxide (SO2). Range 0 to 150ppm, Resolution 0.1ppm. Response time 5 seconds

£509.00

Need help finding the right product?

Description

Sulphur Dioxide (SO2​) is a colourless gas with a sharp, pungent, irritating odour, often described as smelling like burnt matches. It is a major air pollutant and a significant respiratory irritant, formed primarily from the combustion of sulphur-containing fuels or industrial processes involving sulphur.

Key Properties:

  • Colourless: Invisible to the naked eye.
  • Sharp, Pungent, Irritating Odour: Detectable at very low concentrations, providing a good warning sign.
  • Highly Soluble in Water: When dissolved in water (e.g., in the lungs or in atmospheric moisture), it forms sulphurous acid (H2​SO3​), which is irritating and corrosive.
  • Density: Approximately 2.2 times heavier than air, so it tends to settle in low-lying areas, pits, trenches, and confined spaces.
  • Non-flammable: It does not burn.

 

Replacement Sensor, Part No. 17124975-5, Properties

Gas Detected
(Abbreviation)
Sensor Type Measurement Response time
(Seconds)
Calibration Gas
& Concentration
Range Resolution T50 T90
Sulphur Dioxide
SO2
Electrochemical 0 to 150ppm 0.1ppm 5 20 SO2
10 ppm

 

Toxic Gas Sensor Cross-sensitivity Table

Target Gas Sensor
CO CO
(H2 Low)
H2S SO2 NO2 Cl2 ClO2 HCN HCl PH3 NO H2 NH3
CO 100 100 1 1 0 0 0 0 0 0 0 20 0
H2S 5 5 100 1 -40 -3 -25 10 300 25 10 20 25
SO2 0 5 5 100 0 0 0 40 0 0 -40
NO2 -5 5 -25 -165 100 45 -70 30 0 -10
Cl2 -10 0 -20 -25 10 100 60 -20 6 -20 0 0 -50
ClO2 20 100
HCN 15 50 1 0 0 100 35 1 0 30 5
HCl 3 5 0 2 0 0 100 0 15 0 0
PH3 -100 425 300 100
NO 25 40 -0.2 1 5 -5 100 30 0
H2 22 3 0.08 0.5 0 0 0 0 0 0 0 100 0
NH3 0 0 0 0 0 0 0 0 0 0 0 0 100

The table above reflects the percentage response provided by the sensor (top row) when exposed to a known concentration of the target gas (column 1).
The numbers were measured under these environmental conditions: 20 °C (68 °F), 50% RH and 1 atm.
The specified cross-interference numbers apply to new sensors only and may vary with time as well as from sensor to sensor.
“-” means no data available.
This table is given as a reference only and is subject to change.

 