omni instruments | sensors and data acquisition equipment

Sales & Technical Support

17124975-1

Replacement MX6 Sensor for Carbon Monoxide (CO). Range 0 to 1,500ppm, Resolution 1.0 ppm. Response Time 10 Seconds (T50)

£431.00

Need help finding the right product?

Description

Carbon Monoxide (CO) is a colourless, odourless, tasteless, and non-irritating gas that is highly toxic. It’s often referred to as the “silent killer” because it’s impossible to detect with human senses, and its presence can lead to severe health effects or death without warning.

Key Properties

  • Colourless, Odourless, Tasteless: Undetectable by human senses
  • Highly Toxic: Interferes with the blood’s ability to carry oxygen
  • Slightly Lighter Than Air: It mixes readily with air, but can still accumulate in enclosed spaces if not properly ventilated
  • Flammable: Can ignite and form explosive mixtures in high concentrations

 

Replacement Sensor, Part No. 17124975-1, Properties

Gas Detected
(Abbreviation)
Sensor Type Measurement Response time
(Seconds)
Calibration Gas
& Concentration
Range Resolution T50 T90
Carbon Monoxide
CO
Electrochemical 0 to 1,500 ppm 1.0 ppm 10 20 CO
100 ppm

 

Toxic Gas Sensor Cross-sensitivity Table

Target Gas Sensor
CO CO
(H2 Low)
H2S SO2 NO2 Cl2 ClO2 HCN HCl PH3 NO H2 NH3
CO 100 100 1 1 0 0 0 0 0 0 0 20 0
H2S 5 5 100 1 -40 -3 -25 10 300 25 10 20 25
SO2 0 5 5 100 0 0 0 40 0 0 -40
NO2 -5 5 -25 -165 100 45 -70 30 0 -10
Cl2 -10 0 -20 -25 10 100 60 -20 6 -20 0 0 -50
ClO2 20 100
HCN 15 50 1 0 0 100 35 1 0 30 5
HCl 3 5 0 2 0 0 100 0 15 0 0
PH3 -100 425 300 100
NO 25 40 -0.2 1 5 -5 100 30 0
H2 22 3 0.08 0.5 0 0 0 0 0 0 0 100 0
NH3 0 0 0 0 0 0 0 0 0 0 0 0 100

The table above reflects the percentage response provided by the sensor (top row) when exposed to a known concentration of the target gas (column 1).
The numbers were measured under these environmental conditions: 20 °C (68 °F), 50% RH and 1 atm.
The specified cross-interference numbers apply to new sensors only and may vary with time as well as from sensor to sensor.
“-” means no data available.
This table is given as a reference only and is subject to change.

A combined gas cell for CO and H2S is also available for the MX6 gas detector and this sensor can be found on the link below

https://www.omniinstruments.co.uk/p/17124975-j

 

MX6 Replacement Sensor Cells

The MX6 gas detectors include field replaceable, plug-in sensors. The user can replace any sensor in the field and have the instrument in service after performing a simple field calibration. Up to 5 sensor can be installed, each in one or more specific location.

We are here to help

Need help with your products or have a question for us on our custom systems? Use the form below to send us a message and our experts will be in touch to help.