Description
Carbon Dioxide (CO2) is a colourless, odourless gas that is a natural component of the Earth’s atmosphere. While it is naturally present in the air we breathe at low concentrations (around 0.04% or 400 ppm), at higher concentrations, it can pose a significant health hazard, acting as an asphyxiant by displacing oxygen.
Key Properties:
- Colourless, Odourless, Tasteless: Like carbon monoxide, it cannot be detected by human senses, making it particularly dangerous.
- Non-toxic (at typical atmospheric levels): However, it is an asphyxiant at higher concentrations, meaning it reduces the available oxygen in the air, leading to suffocation.
- Heavier than Air: It tends to settle in low-lying areas, pits, basements, and confined spaces, displacing oxygen.
- Non-flammable: It does not burn and is often used in fire suppression.
Replacement Sensor, Part No. 17124975-Q, Properties
Gas Detected (Abbreviation) |
Sensor Type | Measurement | Response time (Seconds) |
Calibration Gas & Concentration |
||
---|---|---|---|---|---|---|
Range | Resolution | T50 | T90 | |||
Carbon Dioxide (CO2) |
Infrared | 0 to 5% Volume | 0.01% Vol | 10 | 25 | CO2 2.5% Vol |
Toxic Gas Sensor Cross-sensitivity Table
Target Gas | Sensor | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
CO | CO (H2 Low) |
H2S | SO2 | NO2 | Cl2 | ClO2 | HCN | HCl | PH3 | NO | H2 | NH3 | |
CO | 100 | 100 | 1 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 20 | 0 |
H2S | 5 | 5 | 100 | 1 | -40 | -3 | -25 | 10 | 300 | 25 | 10 | 20 | 25 |
SO2 | 0 | 5 | 5 | 100 | 0 | 0 | 0 | – | 40 | – | 0 | 0 | -40 |
NO2 | -5 | 5 | -25 | -165 | 100 | 45 | – | -70 | – | – | 30 | 0 | -10 |
Cl2 | -10 | 0 | -20 | -25 | 10 | 100 | 60 | -20 | 6 | -20 | 0 | 0 | -50 |
ClO2 | – | – | – | – | 20 | 100 | – | – | – | – | – | – | – |
HCN | 15 | – | – | 50 | 1 | 0 | 0 | 100 | 35 | 1 | 0 | 30 | 5 |
HCl | 3 | – | – | 5 | 0 | 2 | 0 | 0 | 100 | 0 | 15 | 0 | 0 |
PH3 | – | – | – | – | – | – | -100 | 425 | 300 | 100 | – | – | – |
NO | 25 | 40 | -0.2 | 1 | 5 | – | – | -5 | – | – | 100 | 30 | 0 |
H2 | 22 | 3 | 0.08 | 0.5 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 100 | 0 |
NH3 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 100 |
The table above reflects the percentage response provided by the sensor (top row) when exposed to a known concentration of the target gas (column 1).
The numbers were measured under these environmental conditions: 20 °C (68 °F), 50% RH and 1 atm.
The specified cross-interference numbers apply to new sensors only and may vary with time as well as from sensor to sensor.
“-” means no data available.
This table is given as a reference only and is subject to change.