Description
The Pressure sensor includes a small venting filter with may become blocked in some environments and may drift by up to 0.3hPa per year. To maintain the most accurate measurements it is recommended that the sensors is checked periodically and replaced every 24 months. A spare sensor can be purchased as a direct replacement unit. An 80 Error Code on the MetConnect relates to issues with the pressure sensor
Specifications
| Range | 600-1100hPa |
| Accuracy | ±0.5hPa Typical (0 to 50°C) |
| Resolution | 0.1hPa |
| Units of Measure | hPa, mBar, mmHg, inHg |
An 80 Error Code on the MetConnect relates to issues with the pressure sensor and may need replacement.