Description
![]() |
|
Number of Input Channels | 20 Channels |
Input Terminal Type | M3 Screw Terminals |
Parameter | Measurement Range | |
Voltages | 20, 50, 100, 200, 500mV, 1, 2, 5, 10, 20, 50, 100V, 1-5V F.S. |
Thermocouple Type | Measuring Temperature Range | Measurement Accuracy | |
R/S | 0 ≤ TS ≤ 100°C | ±4.5°C | |
100 < TS ≤ 300°C | ±3.0°C | ||
R: 300 < TS ≤ 1600°C | ±2.2°C | ||
S: 300 < TS ≤ 1760°C | ±2.2°C | ||
B | 400 ≤ TS ≤ 600°C | ±3.5°C | |
600 < TS ≤ 1820°C | ±2.5°C | ||
K | -200 ≤ TS ≤ -100°C | ±1.5°C | |
-100 < TS ≤ 1370°C | ±0.8°C | ||
E | -200 ≤ TS ≤ -100°C | ±1.0°C | |
-100 < TS ≤ 800°C | ±0.8°C | ||
T | -200 ≤ TS ≤ -100°C | ±1.5°C | |
-100 < TS ≤ 400°C | ±0.6°C | ||
J | -200 ≤ TS ≤ -100°C | ±1.0°C | |
100 < TS ≤ 100°C | ±0.8°C | ||
100 < TS ≤ 1100°C | ±0.6°C | ||
N | -200 ≤ TS ≤ 0°C | ±2.2°C | |
0 < TS ≤ 1300°C | ±1.0°C | ||
C(W) | 0 ≤ TS ≤ 2000°C | ±1.8°C | |
Reference contact compensation accuracy | ±0.3°C |
Resistance Bulb Type | Measuring Temperature Range | Applied Current | Measurement Accuracy |
PT100 | -200 ≤ TS ≤ 100°C | 1mA | ±0.6°C |
100 < TS ≤ 500°C | ±0.8°C | ||
500 < TS ≤ 850°C | ±1.0°C | ||
JPT100 | -200 ≤ TS ≤ 100°C | 1mA | ±0.6°C |
100 < TS ≤ 500°C | ±0.8°C | ||
PT1000 | -200 ≤ TS ≤ 100°C | 0.3mA | ±0.6°C |
100 < TS ≤ 500°C | ±0.6°C |
B-530 Humidity Sensor | Measurement Range | |
Humidity | 0-100% Using optional |