| Thermocouple Type |
Measuring Temperature Range |
Measurement Accuracy |
| R/S |
0 ≤ TS ≤ 100°C |
±5.2°C |
| 100 < TS ≤ 300°C |
±3.0°C |
| R: 300 < TS ≤ 1600°C |
± (0.05% of reading +2.0°C) |
| S: 300 < TS ≤ 1760°C |
± (0.05% of reading +2.0°C) |
| B |
400 ≤ TS ≤ 600°C |
±3.5°C |
| 600 < TS ≤ 1820°C |
± (0.05% of reading +2.0°C) |
| K |
-200 ≤ TS ≤ -100°C |
± (0.05% of reading +2.0°C) |
| -100 < TS ≤ 1370°C |
± (0.05% of reading +1.0°C) |
| E |
-200 ≤ TS ≤ -100°C |
± (0.05% of reading +2.0°C) |
| -100 < TS ≤ 800°C |
± (0.05% of reading +1.0°C) |
| T |
-200 ≤ TS ≤ -100°C |
± (0.1% of reading +1.5°C) |
| -100 < TS ≤ 400°C |
± (0.1% of reading +0.5°C) |
| J |
-200 ≤ TS ≤ -100°C |
±2.7°C |
| 100 < TS ≤ 100°C |
±1.7°C |
| 100 < TS ≤ 1100°C |
± (0.05% of reading +1.0°C) |
| N |
-200 ≤ TS ≤ 0°C |
± (0.1% of reading +2.0°C) |
| 0 < TS ≤ 1300°C |
± (0.1% of reading +1.0°C) |
| C(W) |
0 ≤ TS ≤ 2000°C |
± (0.1% of reading +1.5°C) |
| Reference contact compensation accuracy |
±0.5°C |