Top Tips for successful measurements using optical electrodes
- Use a non-fluorescent container for calibration solutions. Some plastics may fluoresce and interfere with calibration.
- Avoid strong artificial lighting during calibration. The alternating nature of artificial lights (especially fluorescent tubes) can interfere with calibration.
- Use a reasonably deep container for calibration in order to ensure a fast temperature response. The calibration solution should preferably be up to the blue line on the Probe.
- Always keep the measurement chamber and electrode lenses clean.
- Always fit the sleeve and end cap during both calibration and measurement.
- Always allow the readings to settle completely during both calibration and measurement.
- Always try to eliminate air bubbles by agitating the Probe after insertion both during calibration and measurement.
- Always calibrate and zero the electrode as close to your sample temperature as possible.
- Always zero the electrode just prior to use in clean water (bottled still mineral water is ideal).
Calibrating the Electrode
A dilution of 100µg/L needs to be prepared by serial dilution of the 1L bottle containing 0.5g Fluorescein Dye Powder
Serial Dilutions
The FSCEIN-CAL makes a stock solution of 500mg/L and needs to be diluted as follows.
Step 1: Transfer 10ml of the 500mg/L solution to a 1L volumetric flask and top up to 1L with deionised water. Put a lid on the 1L flask and invert 10 times.
This step results in a 1 in 100 dilution of the solution. The concentration of this solution is 5mg/L.
Step 2: Transfer 20ml of the 5mg/L solution to a 1L volumetric flask and top up to 1L with deionised water. Put a lid on the 1L flask and invert 10 times.
This step results in a 1 in 50 dilution of the solution from step 1. The concentration of this solution is 100μg/L. This solution can now be used as Pt-2 calibration of the FSCEIN sensor.
Dilute solutions can be stored in a dark bottle in a refrigerator for up to five days. After that time it must be discarded. Solution should be allowed to return to 20°C for calibration process.
Zero Point Calibration
This calibration should be carried out before each day’s use.
- Step 1: Prepare Calibration Bottle. Fill a calibration bottle with clean water (bottled still mineral water is recommended). Never use sparkling or carbonated water.
- Step 2: Insert Probe. Wash the Probe in clean water, then insert it all the way into the calibration bottle. Ensure the Sleeve End Cap is fitted. Gently agitate the Probe several times to remove any clinging air bubbles.
- Step 3: Power On and Stabilise. Switch the Aquameter, or BlueLink Device, on and wait until the temperature and solutions readings are stable. If readings are high, re-agitate the Probe to remove any air bubbles.
- Step 4: Navigate to Calibration Menu. Press the MENU key, then select “Calibration”. The screen will display calibration options including “ZERO?” and “Pt-2”.
- Step 5: Select Zero Calibration. Select “ZERO”. The Meter will display “PLEASE WAIT Stabilising” and a progress counter.
- Step 6: Complete Calibration. Once the calibration is successful, the counter will reach 100%, and “Calibrating 100% Press [OK]” will be displayed, along with the voltage output in mV. This value is not stored in the memory, so a manual record should be made for future reference. Press OK to continue.
- Step 7: Verify Zero Calibration. After calibrating, remove the Probe from the bottle, reinsert it, agitate, and allow the reading to settle.
Calibrating the High Point (Pt-2)
The probe should be removed from the zero solution and dry outer sleeve with a soft cloth.
- Step 1: Prepare Calibration Solution. Prepare the solution as direct above for the FSCEIN probe and pour 300ml in to a clean calibration cup. Gently lower the Probe into the solution all the way. Agitate the Probe several times to remove air bubbles.
- Step 2: Access Calibration Menu. Press the MENU key, then select “Calibration”.
- Step 3: Select Calibration. Select “Pt-2”. The screen will display “PLEASE WAIT Stabilising” and a progress counter. The Meter will wait for stability, then calibrate.
- Step 4: Complete Calibration. After successful calibration, the “Calibrating 100% Press [OK]” screen will appear with the voltage output in mV. This value is not stored in the memory, so a manual record should be made. Press OK to continue.
Any significant difference in the mV values to previous readings could highlight an issue with the electrode. This may just be a dirty lens, incorrectly prepared solution, change in temperature, air bubbles or changes in the measurement chamber. Repeating the calibration after the probe has been cleaned, may resolve any issues.