PA33-35Xhigh accuracy digitally compensated pressure transducers and transmitters
Series 33/35x high accuracy digitally compensated pressure transducers and transmitters, ranges up to 1000 bar, accuracy up to 0.01% total error band using RS485 serial The 4-20mA O/P has a total error band 0.1% ( linearity, hysterises, temperature effect ), ranges 200mBar to 1000 Bar, gauge and absolute.
Highly Precise (0,01%) Pressure Transmitters Series 33 X / 35 X
Digital Output of Transmitter
This high precision of 0,01 %FS is available as an option (the standard Series 33 X has an accuracy of 0,05 %FS). These Series are based on the stable, floating piezoresisitive transducer and the newly developed XEMICS micro-processor with integrated 16 bit A/D converter. Temperature dependencies and non-linearities of the sensor are mathematically compensated. With the READ30 software and the KELLER cable K-107, the calculated pressure can be displayed on a Palmtop, Laptop or PC. The READ 30 software also allows the recording of pressure signals and the graphic display on the PC. Up to 128 transmitters can be hooked together to a Bus-system.
Transmitter with Analog Output
Integrated in the XEMICS processor is a D/A converter of 16 bit for analog signal outputs of 4…20 mA or 0…10 V. The output rate is 400 Hz. The accuracy is diminished by this converting process by 0,05 %FS. The digital output is available on all transmitters with analog output.
Programming
With the KELLER software READ 30 and PROG30, a RS485 converter (i.e. K102 or K107 from KELLER) and a PC, the pressure can be displayed, the units changed, a new gain or zero set. The analog output can be set to any range within the compensated range.
Specification
Pressure Ranges
0,8...1000 bar (abs./rel.)
Output Signal
RS485 (digital) 4...20 mA (2-wire) 0...10 V (3-wire)
Gauges: 1 mbar or 0,05 %FS Absolute: 0,5 mbar or 0,025 %FS (10…40 °C)
Resolution
0,002 %FS
* Only for Series 33 X and for ranges higher than 10 bar.
Accuracy and Precision “Accuracy” is an absolute term, “Precision” a relative term. Dead weight testers are primary standards for pressure, where the pressure is defined by the primary values of mass, length and time. Highest class primary standards in national laboratories indicate the uncertainty of their pressure references with 70 to 90 ppM or close to 0,01%. Commercial dead weight testers as used in our facilities to calibrate the transmitters indicate an uncertainty or accuracy of 0,025 %. Below these levels, KELLER use the expression “Precision” as the ability of a pressure transmitter to be at each pressure point within 0.01 %FS relative to these commercial standards. The transmitter’s full-scale output can be set up to match any standard of your choice by correcting the gain with the PROG30 software.
For further information,
technical advice, or a
quotation please call or
email
us. If you are not sure which
product you need one of our
technical sales advisors will be
happy to discuss your
application. We also accept
company purchase orders by fax
or post. Credit card orders can
also be placed over the phone
for any product.
ref
description
downloads
price (£)
Order Item
Pressure Transducers
PA33/35X
Series 33/35x high accuracy digitally compensated pressure transducers and transmitters, ranges up to 1000 bar. With standard accuracy analogue output, 4-20mA, 0-10v
Series 33/35x high accuracy digitally compensated pressure transducers and transmitters, ranges up to 1000 bar. with high precision 0.01% RS485 output
520.00
Not Available online
Programming Kit
K107
Series 30 programming kit comprising of: interface unit, data transfer cable, amphenal plug adaptor for series 30 transmitters, mains power lead, Palm programming lead, and READ30/PROG 30 software.